1. Remove the eyepiece cover (2).
2. Select the Ring Adapter (12) (sold separately) with the insert (11) of the desired diameter depending on the outer diameter of the lens of your optical device (10) (see Table). The designation 42 mm / 50 mm / 56 mm in the name of the adapter means the lens diameter of the optical device.
3. Screw together the Ring Adapter (12) and the thermal imaging module along the threads of the mounting area (17) until it stops. Then untighten a little (no more than one turn) so that the lever (16) is on the right side (see Figure).
4. Evenly tighten the screws (13) until the ball joint grips in the Ring Adapter (12).
5. Apply 2-3 strips of double-sided tape to the outer surface of the insert (11) of your choice.
6. Push the insert (11) of your choice into the Ring Adapter (12) until it stops.
7. Move the lever (16) to the OPEN position.
8. Before installing the Ring Adapter (12) onto the optical device, it is recommended to degrease the lens body of the optical device (10).
9. Mount the Ring Adapter (12) with the insert (11) onto the lens of the daylight optical device (10) as far as it will go.
10. If the Ring Adaptor (12) with the insert (11) selected according to the table cannot be mounted onto the lens (10), follow the steps below:
11. Move the lever (16) from its initial OPEN position to the CLOSE position.
12. Loosen the locking screw (14) with a 2mm Allen key, if it hasn’t been done before.
13. Tighten the screw (15) using a 4mm Allen key. The clamping force should be 1.5-2 Nm (use a torque screwdriver) to ensure the lever is correctly tightened (16), while the Ring Adapter with the thermal imaging module should not move relative to the body of the optical device (10). If necessary, tighten or loosen the screw (15) to operate the lever (16) in the best way possible.
14. Tighten the locking screw (14) as far as it will go.
15. Turn on the thermal imaging module by briefly pressing the ON/OFF (6) button.
17. Turn on the device and align the image center on the Proton XQ30 display with the image center of the daylight optical device by carefully tilting the thermal imaging module.
18. Align the top and bottom display boundaries of the Proton XQ30 parallel to the horizontal line of the daylight optical device’s reticle. Having reached the best possible position of the thermal imaging module, tighten the two screws (13) until stop. The clamping force should be 6.5-7.5 N·m (use a torque screwdriver to check).
Selection table for optical device inserts
Ring Adapter model |
The internal diameter of the insert needs to match the outer diameter of the objective lens housing of the daylight optical device it is being installed on. |
|
Insert internal diameter, mm |
Suitable for lens housing of daylight optical devices with an outer diameter of, mm |
|
PSP Ring Adapter 42 mm |
45.5 |
45.5 |
46 |
46 |
|
46.5 |
46.5 |
|
47 |
46.7-47.6 |
|
48 |
47.7- 48.6 |
|
49 |
48.7-49.6 |
|
50 |
49.7-50.6 |
|
PSP Ring Adapter 50 mm |
51.6 |
51.6 |
53.4 |
53.4 |
|
55 |
54.7-55.6 |
|
56 |
55.7-56.6 |
|
57 |
56.7-57.6 |
|
58 |
57.7-58.6 |
|
59 |
58.7-59.6 |
|
PSP Ring Adapter 56 mm |
60 |
59.7-60.6 |
61 |
60.7-61.6 |
|
62 |
61.7-62.6 |
|
63 |
62.7-63.6 |
|
64 |
63.7-64.6 |
|
65 |
64.7-65.6 |